Original contribution
Splatter during jet irrigation cleansing of a wound model: A comparison of three inexpensive devices

https://doi.org/10.1016/S0196-0644(05)81259-4Get rights and content

Study objective:

Pressurized jet irrigation is commonly used to cleanse traumatic wounds but results in splatter of blood, a biohazard. Three inexpensive irrigation devices were compared to assess the degree of splatter produced: a 1.25-in. 18-gauge angiocath, an Irrijet Irrigation System with a 12.7-cm splash shield, and a Zerowet Splashield held directly against the wound (Zerowet-C) and held 4 to 10 cm from the wound, an incorrect technique (Zerowet-I).

Design:

A standard laceration was created in pieces of beef. This wound model was placed 1 m from the floor. Paper grid sheets were placed on the irrigator's face and chest. Six grid sheets were suspended at the 9:00, 12:00, and 3:00 positions 1 m from the wound model and 1 and 1.5 m from the floor to simulate exposure to nearby individuals. Two grid sheets were placed flat on the floor, at the 10:30 and 1:30 positions, 1 m from the base of the wound model stand. The study area was contained in a 3 × 2 × 2 m plastic sheet enclosure to prevent air drafts.

Intervention:

Ten irrigations were performed with the angiocath, Irrijet, Zerowet-I, and Zerowet-C. Each run used 200 mL methylene blue solution delivered with a 50-mL syringe by one-hand pressure. The methylene blue splatter on each of the grids was counted by size (diameter, less than 1 mm, more than 1 mm and less than 5 mm, more than 5 mm and less than 10 mm, and more than 10 mm).

Results:

There was significantly less splatter onto the irrigator's face and chest with Irrijet, Zerowet-I, and Zerowet-C. No facial splatter occurred with Zerowet-C. There was significantly less splatter at the 9:00 and 12:00 positions at both heights, and on the floor with Irrijet, Zerowet-I, and Zerowet-C. Less significant splatter difference was noted at the 3:00 position.

Conclusion:

Irrijet, Zerowet-I, and Zerowet-C were superior to the angiocath in preventing splatter during this wound model irrigation. The correct of Zerowet (Zerowet-C) was particularly effective in preventing splatter onto the irrigator's face.

References (7)

There are more references available in the full text version of this article.

Cited by (0)

Presented at the Society for Academic Emergency Medicine Annual Meeting in Toronto, Ontario, Canada, May 1992.